The settings below represent the standards set at the beginning of the project for E-Series data collection. There may be a small proportion of studies and signals that do not match these standards exactly. Please review the settings at the individual sleep study level as you proceed with any analyses.
Channel | EDF Label | Input 1 | Input 2 | Sampling rate (Hz) | Hardware filters (Hz) | Sensor type |
---|---|---|---|---|---|---|
Left Frontal EEG | F3 | F3 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Gold cup electrode |
Right Frontal EEG | F4 | F4 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Gold cup electrode |
Left Central EEG | C3 | C3 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Gold cup electrode |
Right Central EEG | C4 | C4 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Gold cup electrode |
Left Occipital EEG | O1 | O1 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Gold cup electrode |
Right Occipital EEG | O2 | O2 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Gold cup electrode |
Left Mastoid EEG | M1 | M1 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Gold cup electrode |
Right Mastoid EEG | M2 | M2 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Gold cup electrode |
Left EOG | E1 | E1 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Gold cup electrode |
Right EOG | E2 | E2 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Gold cup electrode |
ECG1 | ECG1 | ECG1 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Ag/AgCl patch |
ECG2 | ECG2 | ECG2 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Ag/AgCl patch |
Left Leg EMG | LLeg1 | LLeg1 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Compumedics piezo leg leads |
Left Leg EMG | LLeg2 | LLeg2 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Compumedics piezo leg leads |
Right Leg EMG | RLeg1 | RLeg1 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Compumedics piezo leg leads |
Right Leg EMG | RLeg2 | RLeg2 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Compumedics piezo leg leads |
Center Chin EMG | Chin1 | Chin1 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Gold cup electrode |
Left Submentalis EMG | Chin2 | Chin2 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Gold cup electrode |
Right Submentalis EMG | Chin3 | Chin3 | Fpz | 256 | High Pass 0.16/Low Pass 105 | Gold cup electrode |
Airflow | Airflow | Airflow | - | 32 | - | Compumedics thermistor |
Abdomen | Abdo | Abdo | - | 32 | - | Compumedics Summit IP RIP |
Thorax | Thorax | Thorax | - | 32 | - | Compumedics Summit IP RIP |
Snore | Snore | Snore | - | 256 | - | Compumedics snore mic |
Sum | Sum | Sum | - | 32 | - | Compumedics Summit IP RIP |
Position | PosSensor | PosSensor | - | 16 | - | Compumedics external position sensor |
Oximetry Status | Ox Status | Ox Status | - | 16 | - | Nonin 8000 sensor |
Pulse | Pulse | Pulse | - | 16 | - | Nonin 8000 sensor |
SpO2 | SpO2 | SpO2 | 16 | - | Nonin 8000 sensor | |
Nasal Pressure | Nasal Pressure | Nasal Pressure | - | 32 | - | Nasal cannula |
CPAP Flow | CPAP Flow | CPAP Flow | - | 32 | - | - |
CPAP Pressure | CPAP Press | CPAP Press | - | 16 | - | - |
Plethysmography | Pleth | Pleth | - | 256 | - | Nonin 8000 sensor |
Derived Heart Rate | Derived HR | Derived HR | - | 64/512 | - | Derived from ECG1 and ECG2 |
Light | Light | Light | - | 64 | - | - |
Manual Pos | Manual Pos | Manual Pos | - | 64 | - | - |
Respiratory Rate | Respiratory Rate | Respiratory Rate | - | 512 | - | - |
REF | - | Fpz | - | - | - | - |
Please reach out to us at support@sleepdata.org or in the Forum if you have questions.