6.5.6 CHAT Standardized Recording Montage

  • The use of this recording montage requires the use of CHAT standardized sensors and placement.
  • The signals to collect do not have to be in any specific order but all must be present on recording.
  • All EOG, EEG signals must be collected to reference. When possible, Chin EMG, Leg EMG and ECG signals will be collected to reference.
  • Sampling rates for signal collection are minimum rates and may be exceeded depending on rates offered by the acquisition system. Sampling rates should be the same among like channels (i.e.: EEG, EMG, Respiratory signals).
  • Signals will be collected unfiltered (at hardware settings). This may vary according to the acquisition system.
  • Peak-to-Peak Range of Amplifier may not be adjustable, depending on acquisition system but must allow for adequate display of the signal.
  • 60 Hz filter may be on for collection
  • Established names for respiratory signals may be preserved if the laboratory provides clear indication of its meaning
Signals (Type) to Collect Name of signal Minimum Sampling Rate (Hz) (may be exceeded) Amplifier Peak-to-Peak (peak-peak must fit the expected size of signal) Amplifier Polarity for Collection
Reference REF 200
Ground 200
L eye movement E1 200 1 MV to 6 MV +↑
R eye movement E2 200 1 MV to 6 MV +↑
F3 F3 200 1 MV to 6 MV +↑
F4 F4 200 1 MV to 6 MV +↑
C3 C3 200 1 MV to 6 MV +↑
C4 C4 200 1 MV to 6 MV +↑
T3 T3 200 1 MV to 6 MV +↑
T4 T4 200 1 MV to 6 MV +↑
O1 O1 200 1 MV to 6 MV +↑
O2 O2 200 1 MV to 6 MV +↑
M1 M1 200 1 MV to 6 MV +↑
M2 M2 200 1 MV to 6 MV +↑
L Chin EMG Lchin 200 1 MV to 6 MV +↑
R Chin EMG Rchin 200 1 MV to 6 MV +↑
C Chin EMG Cchin 200 1 MV to 6 MV +↑
ECG1 ECG1 200 1 MV to 6 MV +↑
ECG2 ECG2 200 1 MV to 6 MV +↑
ECG3 ECG3 200 1 MV to 6 MV +↑
RLeg1 EMG RLeg1 200 6 MV to 20 MV +↑
RLeg2 EMG RLeg2 200 6 MV to 20 MV +↑
LLeg1 EMG LLeg1 200 6 MV to 20 MV +↑
LLeg2 EMG LLeg2 200 6 MV to 20 MV +↑
Nasal/Oral Airflow Airflow 32 40 uV to 2 MV +↑
Nasal Pressure CannulaFlow 128 20 uV to 2 MV +↑
Thoracic IP Chest 32 20 uV to 10 MV +↑
Abdominal IP ABD 32 20 uV to 10 MV +↑
IP Sum SUM 32 20 uV to 10 MV +↑
Oximetry (SpO2) Plethysomnography wave form SAO2 Pleth (or other quality indicator) 10 or highest allowable rate Set (0 to 1 V)
Position Position 1.0-10 Set (500 MV to5V)
Capnography waveform numerical EtCO2-W 32 Set (0 to 1 V)
Capnography waveform numerical EtCO2# 15 Set (0 to 1 V)
The following signal is not obligate for CHAT. Collect if done clinically.
Snore Mic Snore 200 24 uV to 1 MV +↑