Signals (Type) to Collect | Name of signal | Minimum Sampling Rate (Hz) (may be exceeded) | Amplifier Peak-to-Peak (peak-peak must fit the expected size of signal) | Amplifier Polarity for Collection |
---|---|---|---|---|
Reference | REF | 200 | ||
Ground | 200 | |||
L eye movement | E1 | 200 | 1 MV to 6 MV | +↑ |
R eye movement | E2 | 200 | 1 MV to 6 MV | +↑ |
F3 | F3 | 200 | 1 MV to 6 MV | +↑ |
F4 | F4 | 200 | 1 MV to 6 MV | +↑ |
C3 | C3 | 200 | 1 MV to 6 MV | +↑ |
C4 | C4 | 200 | 1 MV to 6 MV | +↑ |
T3 | T3 | 200 | 1 MV to 6 MV | +↑ |
T4 | T4 | 200 | 1 MV to 6 MV | +↑ |
O1 | O1 | 200 | 1 MV to 6 MV | +↑ |
O2 | O2 | 200 | 1 MV to 6 MV | +↑ |
M1 | M1 | 200 | 1 MV to 6 MV | +↑ |
M2 | M2 | 200 | 1 MV to 6 MV | +↑ |
L Chin EMG | Lchin | 200 | 1 MV to 6 MV | +↑ |
R Chin EMG | Rchin | 200 | 1 MV to 6 MV | +↑ |
C Chin EMG | Cchin | 200 | 1 MV to 6 MV | +↑ |
ECG1 | ECG1 | 200 | 1 MV to 6 MV | +↑ |
ECG2 | ECG2 | 200 | 1 MV to 6 MV | +↑ |
ECG3 | ECG3 | 200 | 1 MV to 6 MV | +↑ |
RLeg1 EMG | RLeg1 | 200 | 6 MV to 20 MV | +↑ |
RLeg2 EMG | RLeg2 | 200 | 6 MV to 20 MV | +↑ |
LLeg1 EMG | LLeg1 | 200 | 6 MV to 20 MV | +↑ |
LLeg2 EMG | LLeg2 | 200 | 6 MV to 20 MV | +↑ |
Nasal/Oral Airflow | Airflow | 32 | 40 uV to 2 MV | +↑ |
Nasal Pressure | CannulaFlow | 128 | 20 uV to 2 MV | +↑ |
Thoracic IP | Chest | 32 | 20 uV to 10 MV | +↑ |
Abdominal IP | ABD | 32 | 20 uV to 10 MV | +↑ |
IP Sum | SUM | 32 | 20 uV to 10 MV | +↑ |
Oximetry (SpO2) Plethysomnography wave form | SAO2 Pleth (or other quality indicator) | 10 or highest allowable rate | Set (0 to 1 V) | |
Position | Position | 1.0-10 | Set (500 MV to5V) | |
Capnography waveform numerical | EtCO2-W | 32 | Set (0 to 1 V) | |
Capnography waveform numerical | EtCO2# | 15 | Set (0 to 1 V) | |
The following signal is not obligate for CHAT. Collect if done clinically. | ||||
Snore Mic | Snore | 200 | 24 uV to 1 MV | +↑ |