Childhood Adenotonsillectomy Trial

3.1.2 Display Montage

  • Post acquisition filters are used for display screen settings.
  • Because of the variability of peak-to-peak settings between equipment the display sensitivity should be set so that signals are clear and desired waveforms can be seen.
  • If a 60 Hz filter was not on for acquisition it may be used on the display montage.
  • EOGs will be displayed to the reference electrode in order to check the signal quality of the reference electrode as well as to map eye movements.
  • Sites may choose to employ split screen to display leg EMG, respiratory channels and SpO2 signal at a different time base (2 or 5 min.) if the acquisition system allows. If this is not possible, all signals will be displayed on 30 sec time base.

Table 2: Display Montage 1 (REF is located at FPz); CHAT Calibration

Channel Name Low Filter (Hz) High Filter (Hz) Display Polarity Display Amplitude
E1-REF 0.3 35 Negative ↑ *
E2-REF 0.3 35 Negative ↑ *
Lchin-REF 10 70 Negative ↑ *
Rchin-REF 10 70 Negative ↑ *
Cchin-REF 10 70 Negative ↑ *
C3-REF 0.3 35 Negative ↑ *
C4-REF 0.3 35 Negative ↑ *
O1-REF 0.3 35 Negative ↑ *
O2-REF 0.3 35 Negative ↑ *
F3-REF 0.3 35 Negative ↑ *
F4-REF 0.3 35 Negative ↑ *
T3-REF 0.3 35 Negative ↑ *
T4-REF 0.3 35 Negative ↑ *
M1-REF 0.3 35 Negative ↑ *
M2-REF 0.3 35 Negative ↑ *
ECG1-REF 1.0 35 Positive ↑ *
ECG2-REF 1.0 35 Positive ↑ *
ECG3-REF 1.0 35 Positive ↑ *
LLeg1-REF 10 70 Negative ↑ *
LLeg2-REF 10 70 Negative ↑ *
RLeg1-REF 10 70 Negative ↑ *
RLeg2-REF 10 70 Negative ↑ *
Airflow 1.0 15 Positive ↑ *
CannulaFlow 0.05 15 Positive ↑ *
SUM 0.05 15 Positive ↑ *
Chest 0.05 15 Positive ↑ *
ABD 0.05 15 Positive ↑ *

* Set Display Amplitude to adequately visualize all waveforms

Display Montage 2

  • EOGs will be displayed to the reference electrode in order to check the signal quality of the reference electrode as well as to map eye movements.

Table 2: Display Montage 2 CHAT PSG

Signal Type Signal Inputs Channel Name Low Filter (Hz) High Filter (Hz) Display Polarity Display Amplitude
EOG E1-REF E1 0.3 35 Negative ↑ *
EOG E2-REF E2 0.3 35 Negative ↑ *
Chin EMG Lchin-Cchin Chin 10 70 Negative ↑ *
EEG F4-M1 F4-M1 0.3 35 Negative ↑ *
EEG C4-M1 C4-M1 0.3 35 Negative ↑ *
EEG O2-M1 O2-M1 0.3 35 Negative ↑ *
EEG T4-M1 T4-M1 0.3 35 Negative ↑ *
EEG F3-M2 F3-M2 0.3 35 Negative ↑ *
EEG C3-M2 C3-M2 0.3 35 Negative ↑ *
EEG O1-M2 O1-M2 0.3 35 Negative ↑ *
EEG T3-M2 T3-M2 0.3 35 Negative ↑ *
ECG ECG3-ECG1 ECG 1.0 35 Positive ↑ *
Oximetry-Pulse Internal DC Pulse *
Oximetry Plethysmography Waveform or other quality indicator Pleth (from Oximetry) or other quality indicator Pleth or name of quality indicator *
Limb EMG LLeg1-LLeg2 LLeg 10 70 Negative ↑ *
Limb EMG RLeg1-RLeg2 RLeg 10 70 Negative ↑ *
Snore Mic (if used) Bipolar Snore 20 100 Positive ↑ *
Nasal/Oral Flow Bipolar Airflow 1.0 15 Positive ↑ *
Nasal Pressure Pressure Transducer CannulaFlow 0.05 15 Positive ↑ *
IP SUM Bipolar SUM 0.05 15 Positive ↑ *
IP Chest Effort Bipolar Chest 0.05 15 Positive ↑ *
IP Abdominal Effort Bipolar ABD 0.05 15 Positive ↑ *
Oximetry numerical Internal DC SAO2 *
capnography Auxiliary DC Cap *
capnography Auxiliary DC EtCO2 *
Body Position Internal DC Position *

* Set Display Amplitude to adequately visualize all signals