17.16.4 Body Position Sensor Calibration Procedures
After the study has been opened in NetBeacon, single click on
View
.
Single click on
Edit
.
Highlight
Physical Channels
on left hand side of the window.
Right click on
Position
toward the bottom of the screen.
Choose
Modify Calibration
from the drop down menu. Calibrate Channel window will appear.
Highlight position.
Move actual position sensor into the appropriate position. Sensor is sensitive.
Wait 10-15 seconds for each position before determining what position it has settled upon.
Single click on
Acquire
directly across from the position you are calibrating.
Repeat process for all 4 positions (right, left, back, prone).
Single click on
Set Calibration
.
Message will appear ‘Calibration updated successfully’.
Single click on
Okay
.
After you are done, to get back to the View screen, single click on
Edit
toggle button.
Single click on
View
toggle button. Signals should now appear on your screen.
Double check your calibration by moving the position sensor into all four positions. Check the movement on your ‘View’ screen (bottom of screen).
17.16.3 Masimo Calibration Procedures
Polysomnography
cfs
polysomnography
17-00-polysomnography-toc.md
17-01-materials-needed.md
17-02-psg-training-cert.md
17-03-01-ppt-prep.md
17-03-02-equipment-prep.md
17-04-in-lab-visit.md
17-05-the-setting.md
17-06-sensor-placement.md
17-07-suggested-order-psg-hookup.md
17-08-01-step1.md
17-08-02-step2.md
17-08-03-00-step3.md
17-08-03-01-identify-landmarks.md
17-08-03-02-measure-electrode-sites.md
17-08-03-03-table5-measurement-chart.md
17-08-03-04-determine-cz.md
17-08-03-05-determine-c4.md
17-08-03-06-determine-c3.md
17-08-03-07-determine-o1-o2.md
17-08-03-08-determine-a1-a2.md
17-08-03-09-determine-eog.md
17-08-03-10-determine-emg.md
17-08-03-11-prep-electrode-sites.md
17-08-03-12-attach-electrodes.md
17-08-03-13-attachment-sites.md
17-08-03-14-techniques-application.md
17-08-04-step4.md
17-08-05-step5.md
17-08-06-step6.md
17-08-07-step7.md
17-08-08-step8.md
17-08-09-step9.md
17-09-01-signal-verification-form.md
17-09-02-impedance-signal-quality-overview.md
17-09-03-impedance-signal-procedure.md
17-09-04-troubleshooting.md
17-09-05-signals-priority.md
17-09-06-sensor-equipment-check.md
17-10-recording-psg.md
17-11-leaving-lab.md
17-12-morning-arrival.md
17-13-00-data-collection.md
17-13-01-automatic-analysis.md
17-13-02-saving-sleep-study.md
17-13-03-opening-study.md
17-13-04-family-ppt-feedback-report.md
17-13-05-reviewing-med-alerts.md
17-13-06-creating-sas-reports.md
17-14-scoring-procedures.md
17-15-00-checking-outliers.md
17-15-01-checking-outliers-procedure.md
17-15-02-troubleshooting.md
17-16-01-electrode-overview.md
17-16-02-clean-disinfect.md
17-16-03-masimo-calibration.md
17-16-04-position-sensor-calibration.md
National Sleep Research Resource
Cleveland Family Study
https://doi.org/10.25822/jmyx-mz90