17.16.4 Body Position Sensor Calibration Procedures

  1. After the study has been opened in NetBeacon, single click on View.
  2. Single click on Edit.
  3. Highlight Physical Channels on left hand side of the window.
  4. Right click on Position toward the bottom of the screen.
  5. Choose Modify Calibration from the drop down menu. Calibrate Channel window will appear.
  6. Highlight position.
  7. Move actual position sensor into the appropriate position. Sensor is sensitive.
  8. Wait 10-15 seconds for each position before determining what position it has settled upon.
  9. Single click on Acquire directly across from the position you are calibrating.
  10. Repeat process for all 4 positions (right, left, back, prone).
  11. Single click on Set Calibration.
  12. Message will appear ‘Calibration updated successfully’.
  13. Single click on Okay.
  14. After you are done, to get back to the View screen, single click on Edit toggle button.
  15. Single click on View toggle button. Signals should now appear on your screen.
  16. Double check your calibration by moving the position sensor into all four positions. Check the movement on your ‘View’ screen (bottom of screen).