The RemLogic Offline Analysis program will be run at the Reading Center prior to review of studies by the Polysomnologist and final scoring to allow generation of reports and visualization of signals and graphic screens. After offline analysis of the record, the study will be reviewed to determine any signal quality issues, possible monitor/electrode malfunctions, pass/fail status and possible Urgent Alert status indicating need for priority scoring.
To be classified reliable sleep data the record must contain at least 4 hours of scorable/ reliable oximetry accompanied by one scorable respiratory channel (either airflow, thoracic, or abdomen) after reported bedtime. Studies containing less than the minimal requirements will not be processed for scoring and the site coordinator will be notified via e-mail to request a repeat study from the participant.